Permanent Home of Project has moved to: https://github.com/tgh-apil/Reusable-N95-Project
Orig: William Ng et al. www.anesthesia.utoronto.ca/faculty/william-ck-ng
Last: 1 May 2020
Please read: Article on Medium
Urgent design, testing, recommendation, upscale in industry, dissemination of stop-gap N95 mask designs and guidelines in light of disrupted N95 supply to COVID-19 era Canada
- This collaborative is an open-source, data-sharing, parallel development of the common goal (see CERN 2.0 License)
- Investigate back-up stop-gap plan for N95 depletion in Ontario, Canada, see Article on Medium.com
- Reuse of N95 is preferred at this stage by decontamination (VPHP, UVGI, Heat).
- github ~/Design Types, or
- dropbox repo (and other big files, please ask William Ng)
- Team chat channel, email William Ng
Current focus:
- form-fit test Simple Silicone Respirator with Intersurgical AirGuard
HMEbacterial/viral filters focus on refining form-fit test APIL NIH Unibody Silicone Respirator with HME bacterial/viral filterspause on this design
Weekly focus:
- Volunteer testing: aim 3 iterations max for every 52 volunteers. (Andrew et al.)
- 2nd Leap: Dual flat filter in tightly-sealed cartridge system to increased surface area for airflow. We better start NOW. (Lakehead-NOSM Team Andre, Mike, Mohammed; UofT R&D Team Andy, Arash, Jessica, Nasa, Natasha; also thanks Jake Bender from GLIA for bayonet designs)
Medium term challenge:
- Industry partner(s) to disseminate and upscale option(s).
- Medical Grade filter sourcing, preferred non-HME as they produce condensation on wearer-side
- Merv-15 x 2-ply, Billings MERV 14 2-Ply, Woodbridge INOAC 5-ply +/- 1-ply Halyard H100 Okay
- Estimation of unit Cost and Time to producing first 1000 units.
- https://3dprint.nih.gov/collections/covid-19-response
- many variations to choose from, apart form the above, which have gained traction and use in several hospitals
- UHN APIL modification to Unibody Silicone with adapter for 15mm to 22mm adapter (TBC).
- created by Christian Petropolis (Plastics Surgery) at HSC Manitoba; UHN APIL refining, validating, developing
- silicone material dragon skin sx pro to 10, durometer 2A - 10A
- attaches to commerical filters 15mm to 22mm; M size fits small to large; planning for S and L sizes.
- original cad, Christian's Github
- Modification of Montana with different sizing, filter size, geometry, nose-pinching mechanism
- working in parallel with USASK on sealing with silicone rims to flexible PLA/TPU mask bodies
- fit-testing done by OSH at KGH
- Flexible TPU with silicone ring applied to mask-rim, large mask filter casing
- website: Instagram #MakeTheMask, https://www.makethemasks.com
- https://blog.cuw.edu/COVIDmasks/, both modification of Montana and dual-cartridge, folded 3DP models.
Prototype, Refinement (Josh et al.)
- Montana S, M, L size (modified) - (CIGITI|Brandon)
- Kingston XS, S, M, L, XL, Snowcone S, M - Fit-test Kingston Masks from Queen's - 23 April 2020
- Manitoba SSR M (S to come) - required 30 April 2020
- Seal and fit problem - seal adequate 24 April 2020, comfort level
- Onepiece Silicone Iteration - required 24 April 2020 - done
- Filtration area needs to be maximally increased for WOB and any remote chance of getting NIOSH testing
- Considering 2 filter cartridge system with tight seal (Lakehead|Andre, APIL|UofT R&D, NOSM|Michael, Jay Pooley)
NIH S, M. (XS needs redesigning)- pause on this 15 April 2020I3D XS, S, M prototypes built- archivedAdapter to ROXON filter functional as stop-gap option(no longer pursued given poor NIOSH testing)single filter cartridge system with N95 medium(no longer pursued given poor breathability with good mask seal)
- James Scott and Jeff Brook from UofT OS & EH 13 April 2020 - proof of concept testing
- excluded single FLAT N95 filter option, either double cartridge system, or pleated bacterial/viral filter required.
- Consider pursuing N90 grade with Merv 15 filter (suggested by Queen's https://www.mcmaster.com/22905k28)
- Portacount or AccuFIT fit-testing as per OSHA/CSA protocol, standard 7 Fit-Factor Rubric: https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.134AppA
- Pooling Non-UHN partners SickKids, Queen's University (KHSC), other OHS to perform CSA standard mask-fit test
- Functional Test as for any in-person testing using "spray under the hood" Bitrex
- UHN OHS involved, Craig Parsons and Sharon Martin (Mgr Frank Turner). UHN IPAC represented by Dr Alon Vaisman.
Other centres through Health Canada
- https://www.intertek.com/canada/ (https://www.intertek.com/scientific-regulatory-consultancy/)
- https://gcttg.com/en/ (textiles transportation geosynthetics)
- http://www.cambridgematerials.com (raw materials)
- Designs ideally be washable in chlorox 1:10 (as for faceshields) and dried
- Will eventually need microbiology to chip in
- Viral testing is a possibility
- TBC: APIL, UofT AScE|MD Team, GLIA
- Mould (Industry)
- Silicone vs. Rubber (Industry)
- Straps (Vahid)
- Filter Material Sewing (Matt)
- HC MDEL, and
- HC COVID-19 PPE
- UHN|APIL
- 3D GTHA
- PPE Kingston
- Partners
- submitted: 12:30pm, 9 April 2020 by Will; responded to Ministry invitation, 18:15 17 April 2020.
- margaret.kinyanjui@uhnresearch.ca with internal template and actual application
- submitted: 25 April 2020 by Will.
- submitted: 12 April 2020 by Carla
- see contacts list
- UHN Peter Munk Innovation Fund: due 20 Apr 2020
- Li Zhang (GLIA volunteer) to raise awareness and medium-term community funding
- William Ng (Lead, Anesthesia, Principal Investigator @ UHN|APIL)
- Azad Mashari (Director of UHN APIL, Anesthesia)
- Andrew Syrett (Testing, Writer, Anesthesia)
- Matt Ratto (TEDA Director)
- Eric Greenwood (SickKids Advisor, Director of Equipment)
- Devin Singh (Testing, Design input, Engineering PhD, Anesthesia)
- James Scott (OS & EH)
- Alon Vaisman (UHN IPAC physician, UHN COVID-19 PPE team, UofT Infectious Diseases)
- Dmitry Rozenberg (Respirology and Lung Transplant)
- Carla Luzzi (Anesthesia)
- Binu Jacobs (Research Scientist and Associate)
- UHN Advanced Perioperative Imaging Lab (https://apil.ca)
- Joshua Hiansen (Design Engineering, Print Lead)
- Vahid Anwari (Production, Biomedical Engineer)
- Nasa Nguyen (Project Manager), email
- Arash Nourimand (R&D Team: CAD-Fusion)
- Natasha Valenton (R&D Team: CAD-Fusion)
CIGITI, SickKids - https://www.cigiti.ca
UofT MD Program | Hamilton SoM "3D-PPE-GTHA"
- Student Body, Community Presence, Logistics
- Sinthu Senthillmohan, Yousuf Ahmed, et al. (Point of Contact)
- Jessica Trac, et al. (R&D Team Coordinator)
- Yu Zou (Materials Sciences, Polymer Printing)
- Chandra Singh (Fluid Dynamics Engineering)
Lakehead|NOSM's Parallel Initiative
- Andre LaFreniere (Engineering, Lakehead)
- Chris Murray (Engineering, Lakehead)
- Siamak Elyasi (Engineering, Lakehead)
- Michael Poling (3D print, face-shields, NOSM Faculty, Division of Clinical Sciences)
- Alannah MaClean (Student Lead, NOSM)
- Victoria McKinnon (MD Program, Inksmith)
- Samveg Shah (MD Program, Engineering)
- Lawrence Porter (Sheridan College, School of Engineering)
- Cesia Quintero & Megan Singh & Matt Snow (Founders)
- Taylor Sweet (YGK 3D, co-designer)
- Grant Tingstad (Engineer)
- Erik Olson (Engineer)
- Steve Cory (Objex Unlimited, CEO)
- James Garel Jones (VertiGrow, co-designer)
GLIA NFP partner - https://glia.org
- Carrie Wakem (Executive Director)
- Jessica Schroeder (Mattermost)
- Jake Bender (LHSC)
Seeking joint grant proposal for Peter Munk Innovation Fund (Aim for 19 April 2020)(CV focus)
- Jason Maynes, SickKids Research Chair (Portacount lab), viral testing lab (TBC)
- offering 3DP machines' spare capacity and engineering help
- Stephane Masse (3DP Engineer)
- Christian Petropolis (3DP Mask Lead, Plastics Surgery)
- We agree to be their Ontario Partner
- Rick Struthers (True North PP, owner)
- Brian Read (ex-plastcs; industry co-ordinator)
- VP (NVH) Martin Mazza - Technical POC regarding filter
- Mukesh Bhargava (Director)
- Sameer Masoud/Masood (Emergency Physician, UHN and Humber)
- Aim to share filtering knowledge, testing results, combine human factors simulation.